Product description, technical parameters and configuration
Performance advantages
Clear display
The large-screen display realizes the Chinese menu-style dialogue, which makes the display more clear and the operation easier. It can display the rising curve and the base current level (the acquisition board needs to be configured), and the temperature setting values and actual values of the oven, sampler, detector, etc. can be displayed at the same time in one screen, which can provide more colorful images. Instrument information.
Advanced microcomputer system, excellent control function
1. High-performance microcomputer temperature control system adopts international advanced manufacturing technology, high temperature control accuracy (better than ± 0.05 °C), high reliability, strong anti-interference ability; 6 independent temperature control zones, highest control Temperature up to 400 °C; limit temperature setting and over-temperature protection function to ensure the safe operation of the instrument.
2、All Chinese keyboard set various control and use parameters (including detector operating parameters), strong logic, easy to operate; instrument with self-diagnosis, power protection, detector settings, range, polarity and current settings and Display and other functions can accurately display the temperature control settings, actual values, retention and analysis time of each way.
High performance large capacity oven
Large capacity oven can accommodate capillary column and double packed column at the same time; oven has rapid heating and rapid cooling, ie automatic rear opening mechanism (from 300°C to 50°C within 10min, can automatically adjust opening angle according to temperature change), and Quasi-room temperature control is achieved, and the oven has a 10th-order, 11-platform program temperature ramp (reverse control software support required).
High-precision double stable gas path
The carrier gas flow path adopts the double stabilizing design of the steady voltage valve and the steady flow valve; the air and hydrogen flow path adopts the stable regulation mode of the pressure-regulating valve plus the needle valve; the steady flow valve and the needle valve adopt the digital scale knob and has high precision. Good repeatability, high reliability, and easy and intuitive flow adjustment.
Flexible sample introduction system to meet various analysis requirements
The instrument can install up to three samplers at the same time. According to the analysis requirements, the best sampler combination can be selected. Each unit can independently control the temperature, and the sampler can be easily assembled and disassembled.
Single Packed Column / Double Packed Column Sampler (PIP):
The column head injection method for packed column heads can be used for a wide range of columns; six-way valves can be added for gas injection analysis; the use of connectors in packed column injectors can easily perform 0.53 wide-bore capillary column analysis.
Single capillary column/double capillary column injector (SPL):
Dedicated capillary column sampler with diaphragm bleed and back pressure valve shunt adjustment for a wide range of capillary columns.
High sensitivity, high stability detector
The instrument can install up to three detectors at the same time, such as FID, TCD detectors and so on.
According to different analysis applications using different combinations, the serial and parallel operation of the detector can be conveniently implemented.
The new FID has a lower detection limit, allowing easy removal and cleaning of the spouts and ion collection components. The precise fixed emitter mounting structure ensures the consistent performance of each instrument.
Unique air insulation structure makes TCD more stable.
High-sensitivity HTCD uses high-resistance thorium tungsten wire with sensitivity up to 10000mV. mL/mg.
New built-in acquisition system
The advanced built-in acquisition device can collect the instrument control status and chromatographic output signals, realize the connection of an external computer with a communication cable, and carry out real-time control and data processing of the instrument. (Need to configure anti-control software and acquisition board)
The real-time control includes various functions such as temperature control of various ways, selection and setting of detectors, temperature increase control of the program, temperature-program curve tracking, and flow value display.
Data processing includes dual-channel (expandable to multi-channel) high-speed chromatographic data acquisition, automatic or manual integration parameter setting, up to 5 quantitative methods, baseline deduction, and report preparation.
Digital carrier gas flow monitoring system
The GC5400 gas system can be equipped with the latest digital flow monitoring unit. The carrier gas flow rate and capillary flow rate value can be displayed on the screen. Up to four groups of flow rates can be displayed, and the flow rate parameters can be stored automatically, facilitating the recording and recall of analysis conditions. Convenient for split flow adjustment and split ratio calculation without the use of a soap film flow meter.
Technical indicators
temperature control
Temperature control range: 7°C~400°C at room temperature (1°C increments)
Temperature control object: oven, front and rear detectors, front and rear injectors, auxiliary, total 6
Cheng Sheng Order: Seven Orders
Cheng Sheng rate: 0.1 °C ~ 40 °C / min (increment 0.1 °C)
Constant temperature time: 0~655min (increment 1min)
Hydrogen flame ionization detector (FID)
Sensitivity: M≤8×10-12g/s Sample C16
The best test results: M ≤ 3 × 10-12g / s sample C16
Noise: ≤5×10-14A
Drift: ≤6×10-13A/h
Linear range: ≥106
Thermal Conductivity Detector (TCD)
Sensitivity:≥5000mV. mL/mg Sample C16
Noise: ≤20μV
Drift: ≤30μV/h
Linear range: ≥104
Standard configuration
Hosts
GC5400 GC mainframe Large screen Chinese monitor, high performance oven, double stable gas path
Sampler
PIP Packed Column Sampler allows for column head injection for all sizes
SPL capillary column sampler with diaphragm purge and back pressure valve shunt adjustment, tail blow adjustment
Detector
FID Hydrogen Flame Ionization Detector With modular structure, convenient for disassembly and cleaning
TCD Thermal Conductivity Detector High Quality Tantalum Tungsten Wire, Constant Current Source Control, Double Packed Column Balance Mode
HTCD High-sensitivity Thermal Conductivity Detector High Quality Tantalum Tungsten Wire, Constant Current Source Control, Double Packed Column Balance Method
Optional accessories
Counter control software and acquisition board Computer control of instrument temperature, lift and data processing
Digital flow display component Displays the flow value through the host screen